Useful numbers start with a defined workflow. Here is what was recorded, what was measured separately and what needs to be confirmed for your application.

Customer field data · review context

Four runs. The complete picture.

These are individual observations, not a guarantee of performance for another configuration.

BioNex field run measurements
RunFormatTransfersTotal timeReported transfer timingCalculated end-to-end rate
14 × 384 deepwell → 1 × 384 deepwell34322:283.84 s916 / hour
2 · filmed4 × 384 deepwell → 1 × 384 deepwell34323:113.76 s888 / hour
34 × 384 deepwell → 1 × 384 deepwell34320:563.57 s983 / hour
45 × 96 well → 1 × 96 deepwell929:034.51 s610 / hour

Source: BioNex project field-run records. End-to-end rates are calculated as transfers ÷ total run duration × 3,600.

Transfer timing and total run duration are different measurements. The reported transfer timing is not total time divided by transfer count. The four reported timings have a median of 3.80 seconds; the filmed run averages about 888 transfers per hour end to end.
Make the comparison useful

Model your workflow,
not an average lab.

Throughput, footprint and capital needs depend on the configuration. Start with your plate formats, methods, queue and required output.

  1. Agree on what counts as a completed sample or plate.
  2. Include handling, storage, consumables and operator time.
  3. Validate the proposed configuration against representative work.